  • 2002 Nov 14 C.
  • Weaverdyck University of Michigan End Plug Conductivity Test Humidity controlled box –3400 V on SC 10M? to ground 4 uncontaminated 4 with dried soap 4 cleaned Dried 1000 Bulles Cleaned 1000 Bulles Uncontaminated Humid Air Input 2002 Nov 14 C.
  • Weaverdyck University of Michigan Leakage Current on Endplugs 1 10 100 1000 10000 100000 98% 98% 98% 97% 88% 90% 79% 89% 84% 69% 65% 60% 57% humidity nanoAmps Raw Endplugs 1000 Bulles Un-cleaned Cleaned of 1000 Bulles 2002 Nov 14 C.
  • Weaverdyck University of Michigan 1000 Bulles and the Fix Technique: Scrub Noryl surfaces with liberal mounts of water using a brush ...